Topics Featured

CLIA Announces New 2024 Updates

April 29, 2024

Topics Featured

Changes to CLIA requirements for Analytical Quality may affect your proficiency testing and quality control programs.

Effective July 11, 2024, these requirement amendments mark the biggest changes in U.S. regulations since 1992 with testing goals tightening across Chemistry, Immunoassay and Hematology.

What does this mean to your proficiency and QC programs?

Proficiency testing metrics will be stricter*:

Test

Previous Standard

2024 Standard

Goal Reduction

Hemoglobin

7%

4%

42.8%

Hematocrit

6%

4%

33.3%

RBC

6%

4%

33.3%

WBC

15%

10%

33.3%

HbA1c**

N/A**

±8%

N/A** 

*Sten Westgard, Westgard Rules, westgard.com

**new analyte reviewed by CLIA in 2024

What does this mean to your proficiency and QC programs?

  • Increased instrument precision to meet the stricter correlation requirements
  • Adjusted QC frequency to address new goals
  • Partnering with manufacturers to ensure that these goals are met

How can Streck help you?

We offer a complete portfolio of quality control materials to give you confidence in the performance of your instrumentation. Learn more here.

For additional information on the 2024 CLIA updates, click the links below:


Maximize your lab’s efficiency with Streck quality controls.

streck
Quality Control

Flow Cytometry: A powerful tool for clinicians

Flow cytometry allows scientists to rapidly analyze thousands of normal cells and identify the presence of rare and abnormal cells. It is based on sim…
Read More
Quality Control

ILQC: How to earn the trust of your clients and their patients

In laboratory medicine, trust is everything. It is estimated that more than 70% of medical decisions are made using laboratory data. For clinicians an…
Read More
Sickle Cell Trait
Quality Control

Athletes must be tested for SCT

Though it has raised alarm in the athletic community, exercising with sickle cell trait (SCT) is generally safe. With proper awareness and education, …
Read More